We are committed to providing high quality analyzing solution to global customers over the world.

EM-5 Gas Analyzer (UV-DOAS)
EM-5 Gas Analyzer (UV-DOAS)
EM-5 Gas analyzer is applied for online gas analysis of industry stack emissions. EM-5 gas analyzer is applied for online gas analysis of industry stack emissions. It is combined with UV Differential Optical Absorption Spectroscopy (UV-DOAS) technology.
The measuring cell adopts the multi-pass optical path technology, featuring low detection limit, high measurement accuracy and fast response. This analyzer can simultaneously measure multiple gas components online, including SO₂, NO, NO₂, O₂ (electrochemical method), H₂S, Cl₂, NH₃ and CH₂O. 
It is widely applied in various industries, including power plant, chemical industry, steel plant, cement plant, glass manufacturing plant, mining industry, bio-mass  boilers, waste incenieration power plant, etc. 
  • EM-5

  • YUAN ZHENG

B2B does not have purchase function
库存: 0
Inquire
Add to Basket

EM-5C gas analyzer (UV-DOAS) is applied for online gas analysis of industry stack emissions. It is combined with  UV Differential Optical Absorption Spectroscopy (UV-DOAS) technology.  

The measuring cell adopts the multi-pass optical path technology, featuring low detection limit, high measurement accuracy and fast response. This analyzer can simultaneously measure multiple gas components online, including SO₂, NO, NO₂, O₂ (electrochemical method), H₂S, Cl₂, NH₃ and CH₂O. 


Features                 

  •      Adopt Ultraviolet Differential Optical Absorption Spectroscopy (UV-DOAS) 

  •      Long optical path measuring chamber (Max. 15m optical path), high sensitivity.

  •      The measuring chamber adopts multiple reflection technology, detection limit as low as 0.1 ppm;

  •      The pulsed xenon light source has more than 10 years service life.

  •      The NO and NO2 can be measured simultaneously, without the converter;

  •      The analysis module has no optical moving parts and is not affected by vibration;

  •      Constant temperature and heating companion, avoid the influence of temperature;

  •      Not affected by background components such as H2O, CH4, CO, NH3, dust, etc

  •      No cross interference between measured gases;

  •      Customizable to meet the OEM and ODM needs for customer.


Application            

Typical application of industry process

  •     Continuous flue gas emission monitoring for power plants (analyze SO2, NOx,  and O2);

  •     DeSOx process monitoring (analyze SO2 and O2)

  •     DeNOx process monitoring (analyze NOx, and O2)

  •     Continuous flue gas emission monitoring of waste incineration (analyze SO2, NOx,  and O2)

  •     Trace Cl2 analysis in PVC process and titanium dioxide production process in chlor-alkali plants (analyze Cl2)

CEMS-5000, EM-5 应用


EM-5  Specification:

Working principle:

UV DOAS (SO2 NO2, NO);  Electrochemical or Zirconia (O2)

Measuring range:

SO2:    0~300/2000/5000 ppm;

NO:      0~500/2000/5000 ppm;

NO2:   0~600/3000 ppm;

O2:       0~25%

measuring range can be customized

Indication error:

≤±2%

Repetitiveness:

≤2%

Zero drift:

≤±2%F.S./7d

Span drift:

≤±2%F.S./7d

Response time:

< 50s (T90)

Relay output:

14x relay output, , DC30V 2A, configurable

Digital input:

6x digital input, configurable

Analog output:

5x 4-20mAoutput, maximum load: 800Ω, configurable

Analog input:

3x 4-20mAoutput, 100Ω, configurable

Communication:

RS485, RS232, ModBus

Power supply:

AC(220±22)V, 200W, (50±1)Hz

Working temperature:

-10°C ~ 40oC

Working humidity:

0~95%RH

Enclosure rating:

IP42

Mounting:19-inch rack mounting

Net weight:

16 Kg

Gross weight:

23 Kg

Dimension:

483mm ×503mm ×133 mm (W×D×H)




EM-5 Model:

EM-5 Analyzer -5气体




EM-5 Measuring Capability:

EM-5 capability